+44 (0)1392 64 00 20
EATON CORPORATION
ISOLATION TRANSFORMER 115/230V MED GRADE C20 INLET 8 C13 OUT

ISOLATION TRANSFORMER 115/230V MED GRADE C20 INLET 8 C13 OUT

Part Number:
IS1800HGDV
Language:
UK

£1,209.16  (Excl VAT)

Introduction

The IS1800HGDV Isolator® Series Dual-Voltage 115/230V 1800W 50/60Hz Medical-Grade™ Transformer with full CAN/ES/IEC/EN 60601-1 listing is ideal for protecting sensitive electronic equipment in patient-care vicinities. It can be configured for either step-up voltage, step-down voltage or pass-through voltage operation. Along with line isolation and continuous EMI/RFI line noise filtering, its surge protection rating of 240 joules suppresses excess voltage to prevent it from damaging connected equipment.

An internal low-impedance isolation transformer component offers 100% isolation from the input mains AC line. Faraday shield reduces the cumulative leakage current of the IS1800HGDV and connected equipment to levels less than 100 microamps, while removing utility switching transients and harmonics generated by other on-site loads and reducing the impact of utility-related surge conditions.

Transformer filtering provides continuous common-mode noise rejection with no wearable parts. The IS1800HGDV reduces 6000V IEEE 587 Category A and B ring wave and combination wave test surges to only 0.5V common mode.

Components connect to eight C13 outlets with fuse overload protection and a lighted on/off switch. The rugged all-metal unit housing ships ready for placement in any industrial, medical, office, network or audio/video environment. Requires a C19 AC input cord with country specific plug (sold separately).

Full CAN/ES/IEC/EN 60601-1 Medical-Grade Listing with C20 Inlet and 8 C13 Outlets
Ideal for protecting sensitive electronic equipment in patient-care vicinities
Reduces cumulative leakage current of Isolator and connected equipment to levels less than 100 µA
Inexpensive alternative to dedicated circuits and site electrical upgrades
Rugged all-metal housing designed to withstand frequent use in busy environments
Requires a C19 AC input cord with country specific plug (sold separately).

Selectable 115/230V Input/Output Voltage Suitable for Most Applications
Step-up, step-down or pass-through voltage operation (input frequency always equals output frequency)
Internal low-impedance toroidal isolation transformer with Faraday shield offers 100% isolation from mains AC power line
Automatically shuts down all outlets in case of overload
Ships ready for placement in any industrial, medical, office, network or audio/video environment

Protects Connected Equipment from Dangerous Spikes, Surges and Line Noise
Line isolation, continuous noise filtering and 240 joules of AC surge protection
Removes EMI/RFI noise, utility switching transients, load-generated harmonics and ground loops
Floating AC output prevents noise coupling from noisy ground circuits
Reduces 6000V IEEE 587 Category A and B ring wave and combination wave test surges to only 0.5V common mode

Features

AC inlet type

C20 coupler

AC outlet types

C13

AC outlets quantity

8 AC outlet(s)

AC input voltage

115 - 230 V

AC input frequency

50 - 60 Hz

Input current (max)

15 A

Short circuit protection

Y

Load capacity

1800 VA

Housing colour

White

Certification

ANSI/AAMI ES60601-1:2005+A2+A1, IEC60601-1 3RD EDITION +A1, cAN/CSA-C22.2 NO. 60601-1:14, CE, NOM

Operational conditions

Operating temperature (T-T)

0 - 40 °C

Storage temperature (T-T)

-15 - 50 °C

Operating relative humidity (H-H)

5 - 95%

Operating altitude

0 - 3000 m

Weight & dimensions

Width

300 mm

Depth

350 mm

Height

140 mm

Weight

23.5 kg

Package width

445 mm

Package depth

500.1 mm

Package height

279.9 mm

Package weight

23.2 kg

Logistics data

Products per master (outer) case

1 pc(s)

Master (outer) case width

445 mm

Master (outer) case length

500.1 mm

Master (outer) case height

279.9 mm

Master (outer) case gross weight

23.2 kg

Country of origin

Taiwan

No options for this product
.